- | === Sample Environment === | + | ====== Sample Environment |
Sample environmemt on PANDA is available and supported mainly from MLZ SE pool. Please find the values of sample space and height you need in the drawings provided by pdf-files. | Sample environmemt on PANDA is available and supported mainly from MLZ SE pool. Please find the values of sample space and height you need in the drawings provided by pdf-files. | ||
- | == * CC Cryostat (liquid-free)== | + | ===== Cryostats ===== |
- | liquid-free | + | ==== CC Cryostat (liquid-free) ==== |
- | {{: | + | liquid-free closed cycle (CC) cryostat is the working-horse on PANDA for experiments above 2.5K and below 320K. To reduce background, a vacuum chamber with large diameter is available. To mount the sample, we need to open the whole cryostat, which means that sample exchange need additional 0.5 - 1 hour for drying the cryostat (depending on the inner temperature before exchange). This cryostat is dedicated to PANDA so it is available all the time and easy to use. |
- | {{: | + | * {{: |
+ | * {{: | ||
- | == * He3 Insert (for liquid-free CC Cryostat)== | + | ==== CCR cryostat ==== |
- | **(exchange by actual drawing with right numbers!!!!)** | + | For reaching the lower (or higher) temperatures than CC cryostat can do, we use CCR top loading cryostats. They can be also booked without a special insert, if user needs quick sample exchange during the experiment. They cannot reach 2.5K as CC cryostat, their minimum temperatures are normally around 3K. |
- | {{: | + | |
- | == * Dilution Insert (for liquid-free CC Cryostat)== | + | |
- | == * He-cooled Variox (possible | + | Following special inserts are available for CCR cryostat from the pool. They need to be booked |
- | {{: | + | === He3 Insert (for CCR) === |
- | {{: | + | He3 insert can reach temperatures from 300K down to 0.35K. Cooling time to base temperature is 3-4hours. Currently running He3 insert is very automatized and is used on PANDA very often. |
+ | * {{: | ||
- | == * 7.5T magnet MAG-V-7.5T | + | === Dilution Insert |
- | http://www.mlz-garching.de/ | + | He3 insert can reach temperatures from 1K down to 80mK. Without running the dilution, it is possible to reach temperatures above 3K. But reaching stable temperatures between 1-3K is not possible. Cooling time to base temperature is roughly 6 hours. Running dilution insert requires serious effort of the sample environment team. |
- | == * 5T magnet JVM1-5.0T (liquid-He cooled, base temp. 1.5K, possible in combination with Kelvinox-diution | + | |
- | http:// | + | ==== He-cooled Variox (possible in combination with Kelvinox-diution insert) ==== |
+ | Description will be added. | ||
+ | * {{: | ||
+ | * {{: | ||
- | == * 15T magnet | + | ===== Magnets ===== |
+ | Description will be added. | ||
+ | ==== 5T magnet | ||
- | {{: | + | room temp. bore for liquid-free CCr Cryostat, possible in combination with 3He/diution inserts |
- | {{: | + | |
- | {{: | + | * Maximum field is only 5T |
+ | * [[http:// | ||
+ | |||
+ | ==== 5T magnet JVM1-5.0T ==== | ||
+ | |||
+ | liquid-He cooled, base temp. 1.5K | ||
+ | |||
+ | * {{: | ||
+ | * [[http:// | ||
+ | |||
+ | ==== 15T magnet ==== | ||
+ | |||
+ | **out of operation** | ||
+ | |||
+ | < | ||
+ | |||
+ | * {{: | ||
+ | | ||
+ | | ||
| |